Accuracy of Dexcom G6 Continuous Glucose Monitoring in Non-Critically Ill Hospitalized Patients With Diabetes

Diabetes Care. 2021 Jul;44(7):1641-1646. doi: 10.2337/dc20-2856. Epub 2021 Jun 7.

Abstract

Objective: Advances in continuous glucose monitoring (CGM) have transformed ambulatory diabetes management. Until recently, inpatient use of CGM has remained investigational, with limited data on its accuracy in the hospital setting.

Research design and methods: To analyze the accuracy of Dexcom G6, we compared retrospective matched-pair CGM and capillary point-of-care (POC) glucose data from three inpatient CGM studies (two interventional and one observational) in general medicine and surgery patients with diabetes treated with insulin. Analysis of accuracy metrics included mean absolute relative difference (MARD), median absolute relative difference (ARD), and proportion of CGM values within 15, 20, and 30% or 15, 20, and 30 mg/dL of POC reference values for blood glucose >100 mg/dL or ≤100 mg/dL, respectively (% 15/15, % 20/20, % 30/30). Clinical reliability was assessed with Clarke error grid (CEG) analyses.

Results: A total of 218 patients were included (96% with type 2 diabetes) with a mean age of 60.6 ± 12 years. The overall MARD (n = 4,067 matched glucose pairs) was 12.8%, and median ARD was 10.1% (interquartile range 4.6, 17.6]. The proportions of readings meeting % 15/15, % 20/20, and % 30/30 criteria were 68.7, 81.7, and 93.8%, respectively. CEG analysis showed 98.7% of all values in zones A and B. MARD and median ARD were higher in the case of hypoglycemia (<70 mg/dL) and severe anemia (hemoglobin <7 g/dL).

Conclusions: Our results indicate that CGM technology is a reliable tool for hospital use and may help improve glucose monitoring in non-critically ill hospitalized patients with diabetes.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Aged
  • Blood Glucose
  • Blood Glucose Self-Monitoring
  • Diabetes Mellitus, Type 1*
  • Diabetes Mellitus, Type 2* / drug therapy
  • Humans
  • Middle Aged
  • Reproducibility of Results
  • Retrospective Studies

Substances

  • Blood Glucose

Associated data

  • figshare/10.2337/figshare.14454357